1. Comparison of UV and charged aerosol detection approach in pharmaceutical analysis of statins.
- Author
-
Nováková L, Lopéz SA, Solichová D, Satínský D, Kulichová B, Horna A, and Solich P
- Subjects
- Aerosols, Atorvastatin, Equipment and Supplies, Heptanoic Acids analysis, Lovastatin analysis, Pyrroles analysis, Simvastatin analysis, Tablets analysis, Chromatography, High Pressure Liquid instrumentation, Hydroxymethylglutaryl-CoA Reductase Inhibitors analysis
- Abstract
CAD (charged aerosol detector) has recently become a new alternative detection system in HPLC. This detection approach was applied in a new HPLC method for the determination of three of the major statins used in clinical treatment-simvastatin, lovastatin and atorvastatin. The method was optimized and the influence of individual parameters on CAD response and sensitivity was carefully studied. Chromatography was performed on a Zorbax Eclipse XDB C18 (4.6 mm x 75 mm, 3.5 microm), using acetonitrile and formic acid 0.1% as mobile phase. The detection was performed using both CAD (20 pA range) and DAD (diode array detector-238 nm) simultaneously connected in series. In terms of linearity, precision and accuracy, the method was validated using tablets containing atorvastatin and simvastatin. The CAD is designated to be a non-linear detector in a wide dynamic range, however, in this application and in the tested concentration range its response was found to be perfectly linear. The limits of quantitation (0.1 microg/ml) were found to be two times lower than those of UV detection.
- Published
- 2009
- Full Text
- View/download PDF