Back to Search
Start Over
Comparison of UV and charged aerosol detection approach in pharmaceutical analysis of statins.
- Source :
-
Talanta [Talanta] 2009 May 15; Vol. 78 (3), pp. 834-9. Date of Electronic Publication: 2009 Jan 15. - Publication Year :
- 2009
-
Abstract
- CAD (charged aerosol detector) has recently become a new alternative detection system in HPLC. This detection approach was applied in a new HPLC method for the determination of three of the major statins used in clinical treatment-simvastatin, lovastatin and atorvastatin. The method was optimized and the influence of individual parameters on CAD response and sensitivity was carefully studied. Chromatography was performed on a Zorbax Eclipse XDB C18 (4.6 mm x 75 mm, 3.5 microm), using acetonitrile and formic acid 0.1% as mobile phase. The detection was performed using both CAD (20 pA range) and DAD (diode array detector-238 nm) simultaneously connected in series. In terms of linearity, precision and accuracy, the method was validated using tablets containing atorvastatin and simvastatin. The CAD is designated to be a non-linear detector in a wide dynamic range, however, in this application and in the tested concentration range its response was found to be perfectly linear. The limits of quantitation (0.1 microg/ml) were found to be two times lower than those of UV detection.
Details
- Language :
- English
- ISSN :
- 1873-3573
- Volume :
- 78
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Talanta
- Publication Type :
- Academic Journal
- Accession number :
- 19269437
- Full Text :
- https://doi.org/10.1016/j.talanta.2008.12.057