1. DQE and eDQE of the KARS-BKS2 Digital Slot Scan X-Ray System
- Author
-
V. A. Labusov, A. V. Bekhterev, R. V. Kotkov, M. S. Khramov, I. I. Strokov, R. A. Lokhtin, and D. A. P’yanov
- Subjects
Physics ,Focal point ,Null (radio) ,business.industry ,Radiation quality ,Biomedical Engineering ,X-ray ,Medicine (miscellaneous) ,Radiation ,Detective quantum efficiency ,Medical Laboratory Technology ,Optics ,Spatial frequency ,business - Abstract
DQE and eDQE were evaluated for a slot scan X-ray system using four standard radiation qualities: RQA3, RQA5, RQA7, and RQA9. The results showed that the system under consideration, KARS-BKS2 fluorograph, provided maximal DQE(u,v) and eDQE(u′,v′) in the case of radiation quality RQA5 and a small focal point. At null spatial frequencies DQE(0) was greater than 40% and eDQE(0) was 35%.
- Published
- 2021
- Full Text