Back to Search Start Over

Technical Note: Simplified and practical pretherapy tumor dosimetry - A feasibility study for 131 I-MIBG therapy of neuroblastoma using 124 I-MIBG PET/CT.

Authors :
Seo, Youngho
Seo, Youngho
Huh, Yoonsuk
Huang, Shih-Ying
Hernandez-Pampaloni, J Miguel
Hawkins, Randall A
Gustafson, W Clay
Vo, Kieuhoa T
Matthay, Katherine K
Seo, Youngho
Seo, Youngho
Huh, Yoonsuk
Huang, Shih-Ying
Hernandez-Pampaloni, J Miguel
Hawkins, Randall A
Gustafson, W Clay
Vo, Kieuhoa T
Matthay, Katherine K
Source :
Medical physics; vol 46, iss 5, 2477-2486; 0094-2405
Publication Year :
2019

Abstract

PurposeRadiation dose calculated on tumors for radiopharmaceutical therapy varies significantly from tumor to tumor and from patient to patient. Accurate estimation of radiation dose requires multiple time point measurements using radionuclide imaging modalities such as SPECT or PET. In this report, we show our technical development of reducing the number of scans needed for reasonable estimation of tumor and normal organ dose in our pretherapy imaging and dosimetry platform of 124 I-metaiodobenzylguanidine (MIBG) positron emission tomography/computed tomography (PET/CT) for 131 I-MIBG therapy of neuroblastoma.MethodsWe analyzed the simplest kinetic data, areas of two-time point data for five patients with neuroblastoma who underwent 3 or 4 times of 124 I-MIBG PET/CT scan prior to 131 I-MIBG therapy. The data for which we derived areas were percent of injected activity (%IA) and standardized uptake value of tumors. These areas were correlated with time-integrated activity coefficients (TIACs) from full data (3 or 4 time points). TIACs are direct correlates with radiation dose as long as the volume and the radionuclide are known.ResultsThe areas of %IAs between data obtained from all the two-time points with time points 1 and 2 (day 0 and day 1), time points 2 and 3 (day 1 and day 2), and time points 1 and 3 (day 0 and day 2) showed reasonable correlation (Pearson's correlation coefficient |r| > 0.5) with not only tumor and organ TIACs but also tumor and organ absorbed doses. The tumor and organ doses calculated using %IA areas of time point 1 and time point 2 were our best fits at about 20% individual percent difference compared to doses calculated using 3 or 4 time points.ConclusionsWe could achieve reasonable accuracy of estimating tumor doses for subsequent radiopharmaceutical therapy using only the two-time point imaging sessions. Images obtained from these time points (within the 48-h after administration of radiopharmaceutical) were also viewed as

Details

Database :
OAIster
Journal :
Medical physics; vol 46, iss 5, 2477-2486; 0094-2405
Notes :
application/pdf, Medical physics vol 46, iss 5, 2477-2486 0094-2405
Publication Type :
Electronic Resource
Accession number :
edsoai.on1391605862
Document Type :
Electronic Resource