Back to Search Start Over

p64 flow diverter: Results in 108 patients from a single center

Authors :
De Beule, T.
Boulanger, T.
Heye, S.
van Rooij, W.J.
van Zwam, W.H.
Stockx, L.
De Beule, T.
Boulanger, T.
Heye, S.
van Rooij, W.J.
van Zwam, W.H.
Stockx, L.
Source :
Interventional Neuroradiology vol.27 (2021) date: 2021-02-01 nr.1 p.51-59 [ISSN 1591-0199]
Publication Year :
2021

Abstract

Background and purposeFlow diverters are increasingly used to treat intracranial aneurysms. We report the safety and efficacy of the p64 flow diverter, a resheathable and detachable device for intracranial aneurysms.Materials and methodsWe retrospectively reviewed 108 patients with 109 aneurysms treated with the p64 between March 2014 and July 2019. There were 87 women and 21 men, mean age 57 years. Of 109 aneurysms, 74 were discovered incidentally, 12 were symptomatic, 18 were previously treated, and five were ruptured dissection aneurysms. A total of 10 aneurysms were located in the posterior circulation. The mean aneurysm or remnant size was 8.1 mm.ResultsHemorrhage by perforation with the distal guidewire occurred in two patients with permanent neurological deficits in one. In one patient, acute in-stent occlusion caused infarction with a permanent deficit. Permanent morbidity was 1.9% (2 of 108, 95%CI 0.1-6.9%); there was no mortality. During follow-up, three in-stent occlusions occurred, all asymptomatic. There were no delayed hemorrhagic complications. At six months, 77 of 96 aneurysms (80.2%) were completely occluded, and at last follow-up, this increased to 93 of 96 aneurysms (96.9%). In-stent stenosis at any degree occurred in 11 patients, progressing to asymptomatic complete occlusion in one. In the other patients, stenosis resolved or improved at further follow-up.ConclusionThe p64 offers an effective and safe treatment option. Aneurysm occlusion rate was 97% at last follow-up, mostly achieved with a single device. There were no delayed hemorrhagic complications. Delayed in-stent stenosis infrequently progresses to occlusion but remains a matter of concern.

Details

Database :
OAIster
Journal :
Interventional Neuroradiology vol.27 (2021) date: 2021-02-01 nr.1 p.51-59 [ISSN 1591-0199]
Notes :
DOI: 10.1177/1591019920932048, English
Publication Type :
Electronic Resource
Accession number :
edsoai.on1390865406
Document Type :
Electronic Resource