Back to Search
Start Over
Clinical and echocardiographic risk factors for device-related thrombus after left atrial appendage closure:an analysis from the multicenter EUROC-DRT registry
- Source :
- Vij , V , Piayda , K , Nelles , D , Gloekler , S , Galea , R , Fürholz , M , Meier , B , Valgimigli , M , O’Hara , G , Arzamendi , D , Agudelo , V , Asmarats , L , Freixa , X , Flores-Umanzor , E , De Backer , O , Sondergaard , L , Nombela-Franco , L , McInerney , A , Korsholm , K , Nielsen-Kudsk , J E , Afzal , S , Zeus , T , Operhalski , F , Schmidt , B , Montalescot , G , Guedeney , P , Iriart , X , Miton , N , Saw , J , Gilhofer , T , Fauchier , L , Veliqi , E , Meincke , F , Petri , N , Nordbeck , P , Ognerubov , D , Merkulov , E , Cruz-González , I , Gonzalez-Ferreiro , R , Bhatt , D L , Laricchia , A , Mangieri , A , Omran , H , Schrickel , J W , Rodes-Cabau , J , Sievert , H , Nickenig , G & Sedaghat , A 2022 , ' Clinical and echocardiographic risk factors for device-related thrombus after left atrial appendage closure : an analysis from the multicenter EUROC-DRT registry ' , Clinical Research in Cardiology , vol. 111 , no. 11 , pp. 1276-1285 .
- Publication Year :
- 2022
-
Abstract
- Background: Data on Device-related Thrombus (DRT) after left atrial appendage closure (LAAC) remain scarce. This study aimed to investigate risk factors for DRT from centers reporting to the EUROC-DRT registry. Methods: We included 537 patients (112 with DRT and 425 without DRT) who had undergone LAAC between 12/2008 and 04/2019. Baseline and implantation characteristics, anti-thrombotic treatment and clinical outcomes were compared between both groups in uni- and multivariate analyses. Additional propensity-score matching (PSM) was conducted to focus on the role of implantation characteristics. Results: Patients with DRT showed higher rates of previous stroke/transient ischemic attack (TIA) (49.1% vs. 34.7%, p < 0.01), spontaneous echocardiographic contrast (SEC) (44.9% vs. 27.7%, p < 0.01) and lower left atrial appendage (LAA) peak emptying velocity (35.4 ± 18.5 vs. 42.4 ± 18.0 cm/s, p = 0.02). Occluders implanted in DRT patients were larger (25.5 ± 3.8 vs. 24.6 ± 3.5 mm, p = 0.03) and implanted deeper in the LAA (mean depth: 7.6 ± 4.7 vs. 5.7 ± 4.7 mm, p < 0.01). Coverage of the appendage ostium was achieved less often in DRT patients (69.5% vs. 81.5%, p < 0.01), while DRT patients were less frequently on oral anticoagulation (7.1% vs. 16.7%, p < 0.01). Multivariate analysis identified age, prior stroke/TIA and SEC as independent risk factors for DRT. After PSM, implantation depth was found to be predictive. Rates of stroke/TIA were higher in DRT patients (13.5% vs. 3.8%, Hazard Ratio: 4.21 [95%-confidence interval: 1.88–9.49], p < 0.01). Conclusions: DRT after LAAC is associated with adverse outcome and appears to be of multifactorial origin, depending on patient characteristics, anticoagulation regimen and device position. Graphical abstract: [Figure not available: see fulltext.].
Details
- Database :
- OAIster
- Journal :
- Vij , V , Piayda , K , Nelles , D , Gloekler , S , Galea , R , Fürholz , M , Meier , B , Valgimigli , M , O’Hara , G , Arzamendi , D , Agudelo , V , Asmarats , L , Freixa , X , Flores-Umanzor , E , De Backer , O , Sondergaard , L , Nombela-Franco , L , McInerney , A , Korsholm , K , Nielsen-Kudsk , J E , Afzal , S , Zeus , T , Operhalski , F , Schmidt , B , Montalescot , G , Guedeney , P , Iriart , X , Miton , N , Saw , J , Gilhofer , T , Fauchier , L , Veliqi , E , Meincke , F , Petri , N , Nordbeck , P , Ognerubov , D , Merkulov , E , Cruz-González , I , Gonzalez-Ferreiro , R , Bhatt , D L , Laricchia , A , Mangieri , A , Omran , H , Schrickel , J W , Rodes-Cabau , J , Sievert , H , Nickenig , G & Sedaghat , A 2022 , ' Clinical and echocardiographic risk factors for device-related thrombus after left atrial appendage closure : an analysis from the multicenter EUROC-DRT registry ' , Clinical Research in Cardiology , vol. 111 , no. 11 , pp. 1276-1285 .
- Notes :
- English
- Publication Type :
- Electronic Resource
- Accession number :
- edsoai.on1382511031
- Document Type :
- Electronic Resource