Acceptable fetal dose using flattening filter-free volumetric arc therapy (FFF VMAT) in postoperative chemoradiotherapy of tongue cancer during pregnancy

Authors :
Takahashi, Wataru
Nawa, Kanabu
Haga, Akihiro
Yamashita, Hideomi
Imae, Toshikazu
Ogita, Mami
Okuma, Kae
Abe, Osamu
Nakagawa, Keiichi
Takahashi, Wataru
Nawa, Kanabu
Haga, Akihiro
Yamashita, Hideomi
Imae, Toshikazu
Ogita, Mami
Okuma, Kae
Abe, Osamu
Nakagawa, Keiichi
Publication Year :
2019

Abstract

Optimizing irradiation protocols for pregnant women is challenging, because there are few cases and a dearth of fetal dosimetry data. We cared for a 36-year-old pregnant woman with tongue cancer. Prior to treatment, we compared three intensity-modulated radiation therapy (IMRT) techniques, including helical tomotherapy, volumetric arc therapy (VMAT), and flattening-filter free VMAT (FFF-VMAT) using treatment planning software. FFF-VMAT achieved the minimum fetal exposure and was selected as the optimal modality. We prescribed 66 Gy to the involved nodes, 60 Gy to the tumor bed and ipsilateral neck, and 54 Gy to the contralateral neck over 33 fractions. To confirm the out-of-field exposure per fraction, surface doses and the rectal dose were measured during FFF-VMAT delivery. Postoperative chemoradiotherapy was delivered using IMRT and a cisplatin regimen. Without any shielding, the total fetal dose was 0.03 Gy, within the limits established by the ICRP. A healthy girl was born vaginally at 37 weeks’ gestation.

Details

Database :
OAIster
Notes :
application/pdf, English
Publication Type :
Electronic Resource
Accession number :
edsoai.on1375223739
Document Type :
Electronic Resource