Back to Search
Start Over
Dual-Energy Computed Tomography for Detection and Characterization of Monosodium Urate, Calcium Pyrophosphate, and Hydroxyapatite:A Phantom Study on Diagnostic Performance
- Source :
- Døssing , A , Müller , F C , Becce , F , Stamp , L , Bliddal , H & Boesen , M 2021 , ' Dual-Energy Computed Tomography for Detection and Characterization of Monosodium Urate, Calcium Pyrophosphate, and Hydroxyapatite : A Phantom Study on Diagnostic Performance ' , Investigative Radiology , vol. 56 , no. 7 , pp. 417-424 .
- Publication Year :
- 2021
-
Abstract
- Objectives The aim of this study was to determine the diagnostic performance of dual-energy computed tomography (DECT) to detect and distinguish crystal deposits in a phantom. The primary objective was to determine the cutoff DECT ratio and the cross-sectional area (CSA) of a crystal deposit necessary to differentiate monosodium urate (MSU), calcium pyrophosphate (CPP), and calcium hydroxyapatite (HA) using DECT. Our secondary objective was to determine the concentration for limit of detection for MSU, CPP, and HA crystal deposits. Exploratory objectives included the comparison between 2 generations of DECT scanners from the same manufacturer as well as different scanner settings. Materials and Methods We used a cylindrical soft tissue phantom with synthetic MSU, CPP, and HA crystals suspended in resin. Crystal suspension concentration increased with similar attenuation between MSU, CPP, and HA in conventional CT. The phantom was scanned on 2 dual-source DECT scanners, at 2 dose levels and all available tube voltage combinations. Both scanners had a tin (Sn) filter at the high-energy spectra. Dual-energy CT ratios were calculated for a given tube voltage combination by dividing linear regression lines of CT numbers against concentration. Dual-energy CT ratios were compared using an analysis of covariance. Receiver operating characteristic curves and corresponding areas under the curve (AUCs) were calculated for individual crystal suspension comparisons (HA vs CPP, MSU vs CPP, and MSU vs HA). Results At standard clinical scan settings with 8 mGy and 80/Sn150 kV, the DECT ratios were as follows: CPP, 2.02 (95% confidence interval [CI], 1.98–2.07); HA, 2.00 (95% CI, 1.96–2.05); and MSU, 1.09 (95% CI, 1.06–1.11). Ratios varied numerically depending on the scanner and tube voltage combination. Monosodium urate crystal DECT ratios were significantly different from HA and CPP (P < 0.001), whereas DECT ratios for HA and CPP crystals did n
Details
- Database :
- OAIster
- Journal :
- Døssing , A , Müller , F C , Becce , F , Stamp , L , Bliddal , H & Boesen , M 2021 , ' Dual-Energy Computed Tomography for Detection and Characterization of Monosodium Urate, Calcium Pyrophosphate, and Hydroxyapatite : A Phantom Study on Diagnostic Performance ' , Investigative Radiology , vol. 56 , no. 7 , pp. 417-424 .
- Notes :
- English
- Publication Type :
- Electronic Resource
- Accession number :
- edsoai.on1322766869
- Document Type :
- Electronic Resource