Back to Search Start Over

Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: In phantom feasibility study

Authors :
Gambarini, G
Carrara, M
Tenconi, C
Mantaut, N
Borroni, M
Cutajar, D
Petasecca, M
Fuduli, I
Lerch, M
Pignoli, E
Rosenfeld, Anatoly B.
Gambarini, G
Carrara, M
Tenconi, C
Mantaut, N
Borroni, M
Cutajar, D
Petasecca, M
Fuduli, I
Lerch, M
Pignoli, E
Rosenfeld, Anatoly B.
Source :
Faculty of Engineering and Information Sciences - Papers: Part A
Publication Year :
2013

Abstract

MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Results show that dual-MOSkin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.

Details

Database :
OAIster
Journal :
Faculty of Engineering and Information Sciences - Papers: Part A
Notes :
application/pdf
Publication Type :
Electronic Resource
Accession number :
edsoai.on1298566678
Document Type :
Electronic Resource