Back to Search Start Over

Risk Factors for Infections Involving Cardiac Implanted Electronic Devices

Authors :
Birnie, D.H.
Wang, Jia
Alings, M.
Philippon, F.
Parkash, R.
Manlucu, J.
Angaran, P.
Rinne, C.
Coutu, B.
Low, R.A.
Essebag, V.
Morillo, C.
Redfearn, D.
Toal, S.
Becker, G.
Degrace, M.
Thibault, B.
Crystal, E.
Tung, S.
LeMaitre, J.
Sultan, O.
Bennett, M.
Bashir, J.
Ayala-Paredes, F.
Gervais, P.
Rioux, L.
Hemels, M.E.W.
Bouwels, L.H.
Exner, D.V.
Dorian, P.
Connolly, S.J.
Longtin, Y.
Krahn, A.D.
Birnie, D.H.
Wang, Jia
Alings, M.
Philippon, F.
Parkash, R.
Manlucu, J.
Angaran, P.
Rinne, C.
Coutu, B.
Low, R.A.
Essebag, V.
Morillo, C.
Redfearn, D.
Toal, S.
Becker, G.
Degrace, M.
Thibault, B.
Crystal, E.
Tung, S.
LeMaitre, J.
Sultan, O.
Bennett, M.
Bashir, J.
Ayala-Paredes, F.
Gervais, P.
Rioux, L.
Hemels, M.E.W.
Bouwels, L.H.
Exner, D.V.
Dorian, P.
Connolly, S.J.
Longtin, Y.
Krahn, A.D.
Source :
Journal of the American College of Cardiology; 2845; 2854; 0735-1097; 23; 74; ~Journal of the American College of Cardiology~2845~2854~~~0735-1097~23~74~~
Publication Year :
2019

Abstract

Item does not contain fulltext<br />BACKGROUND: Cardiac implantable electronic device infection is a major complication that usually requires device removal. PADIT (Prevention of Arrhythmia Device Infection Trial) was a large cluster crossover trial of conventional versus incremental antibiotics. OBJECTIVES: This study sought to investigate independent predictors of device infection in PADIT and develop a novel infection risk score. METHODS: In brief, over 4 6-month periods, 28 centers used either conventional or incremental prophylactic antibiotic treatment in all patients. The primary outcome was hospitalization for device infection within 1 year (blinded endpoint adjudication). Multivariable logistic prediction modeling was used to identify the independent predictors and develop a risk score for device infection. The prediction models were internally validated with bootstrap methods. RESULTS: Device procedures were performed in 19,603 patients, and hospitalization for infection occurred in 177 (0.90%) within 1 year of follow-up. The final prediction model identified 5 independent predictors of device infection (prior procedures [P], age [A], depressed renal function [D], immunocompromised [I], and procedure type [T]) with an optimism-corrected C-statistic of 0.704 (95% confidence interval: 0.660 to 0.744). A PADIT risk score ranging from 0 to 15 points classified patients into low (0 to 4), intermediate (5 to 6) and high (>/=7) risk groups with rates of hospitalization for infection of 0.51%, 1.42%, and 3.41%, respectively. CONCLUSIONS: This study identified 5 independent predictors of device infection and developed a novel infection risk score in the largest cardiac implantable electronic device trial to date, warranting validation in an independent cohort. The 5 independent predictors in the PADIT score are readily adopted into clinical practice. (Prevention of Arrhythmia Device Infection Trial [PADIT Pilot]; NCT01002911).

Details

Database :
OAIster
Journal :
Journal of the American College of Cardiology; 2845; 2854; 0735-1097; 23; 74; ~Journal of the American College of Cardiology~2845~2854~~~0735-1097~23~74~~
Publication Type :
Electronic Resource
Accession number :
edsoai.on1284115119
Document Type :
Electronic Resource