Ion selective electrodes are not suitable for measurements of silver ion concentrations in alkaline carbonate media

Authors :
Hedberg, Jonas
Lowe, Troy A.
Wold, Susanna
Odnevall Wallinder, Inger
Hedberg, Jonas
Lowe, Troy A.
Wold, Susanna
Odnevall Wallinder, Inger
Publication Year :
2013

Abstract

An ion selective electrode (ISE) for determination of the labile silver ion concentration in carbonate containing solutions of pH 10 was seen to give incorrect results due to shifts in the Ag vertical bar Ag+ equilibrium. This drawback was not the case for the differential pulse anodic stripping voltammetry (DPASV) method.<br />QC 20130311