Philips Intellivue NMT module: precision and performance improvements to meet the clinical requirements of neuromuscular block management

Authors :
UCL - SSS/IREC/PNEU - Pôle de Pneumologie, ORL et Dermatologie
UCL - SSS/IREC/MEDA - Pôle de médecine aiguë
UCL - SSS/IREC/MONT - Pôle Mont Godinne
UCL - (MGD) Service d'anesthésiologie
UCL - (MGD) Unité de support scientifique
UCL - (MGD) Service d'oto-rhino-laryngologie
Dubois, Virginie
Fostier, Guillaume
Dutrieux, Marie
Jamart, Jacques
Collet, Stéphanie
De Dorlodot, Clotilde
Eloy, Philippe
Dubois, Philippe E
UCL - SSS/IREC/PNEU - Pôle de Pneumologie, ORL et Dermatologie
UCL - SSS/IREC/MEDA - Pôle de médecine aiguë
UCL - SSS/IREC/MONT - Pôle Mont Godinne
UCL - (MGD) Service d'anesthésiologie
UCL - (MGD) Unité de support scientifique
UCL - (MGD) Service d'oto-rhino-laryngologie
Dubois, Virginie
Fostier, Guillaume
Dutrieux, Marie
Jamart, Jacques
Collet, Stéphanie
De Dorlodot, Clotilde
Eloy, Philippe
Dubois, Philippe E
Source :
Journal of Clinical Monitoring and Computing, Vol. 34, p. 111-116 (2020)
Publication Year :
2020

Abstract

offour (TOF) ratio during neuromuscular block (NMB) recovery. This study evaluated the precision and performance of the Philips Intellivue NMT module (NMT) before (part 1) and after (part 2) several technical upgrades (i.e., firmware upgrade, new cable, and hand adapter) that were recently available. Two cohorts of 30 patients who were scheduled to undergo rhino/ septoplasty under general anesthesia were included in the study. TOF ratios were recorded simultaneously every 15 s on both hands with the NMT and a TOF-Watch SX installed inside a SL TOF-Tube (TWX). Before rocuronium was administered and once final responses were stabilized, the average of the four successive measurements that determined the baselines and repeatability coefficients were compared using a z test. Simultaneous measurements were recorded at different NMB stages: onset, depth of NMB after intubation, when TWX recovered TOF count 2, TOF ratios 0.5 and 0.9, and when NMT recovered TOF ratio 0.9. The results were compared using a Student t test; p < 0.05 was considered significant. The NMT repeatability coefficients obtained in part 1 were significantly higher than with the TWX, they were significantly lower in part 2. Initially, the NMT significantly overestimated NMB recovery at every stage. Conversely, in the second part of the study, no difference reached statistical significance. With the recent upgrades and the new hand adapter, the NMT provided similar results compared with the TWX, Their implementation should be recommended in clinical practice.