Back to Search Start Over

Rib-Cage-Movement Measurements as a Potential New Trigger Signal in Non-Invasive Mechanical Ventilation

Authors :
Ivanović, Marija D.
Petrović, Jovana S.
Miletić, Marjan
Daničić, Aleksandar
Bojović, Boško P.
Vukcevic, M.
Lazovic, B.
Gluvić, Zoran
Hadžievski, Ljupčo
Allsop, Thomas P.
Webb, David J.
Ivanović, Marija D.
Petrović, Jovana S.
Miletić, Marjan
Daničić, Aleksandar
Bojović, Boško P.
Vukcevic, M.
Lazovic, B.
Gluvić, Zoran
Hadžievski, Ljupčo
Allsop, Thomas P.
Webb, David J.
Publication Year :
2015

Abstract

Non-invasive ventilation performed through an oronasal mask is a standard in clinical and homecare mechanical ventilation. Besides all its advantages, inevitable leaks through the mask cause errors in the feedback information provided by the airflow sensor and, hence, patient-ventilator asynchrony with multiple negative consequences. Here we investigate a new way to provide a trigger to the ventilator. The method is based on the measurement of rib cage movement at the onset of inspiration and during breathing by fibre-optic sensors. In a series of simultaneous measurements by a long-period fibre grating sensor and pneumotachograph we provide the statistical evidence of the 200 ms lag of the pneumo with respect the fibre-optic signal. The lag is registered consistently across three independent delay metrics. Further, we discuss exceptions from this trend and identify the needed improvements to the proposed fibre-sensing scheme.

Details

Database :
OAIster
Publication Type :
Electronic Resource
Accession number :
edsoai.on1085026216
Document Type :
Electronic Resource