Rib-Cage-Movement Measurements as a Potential New Trigger Signal in Non-Invasive Mechanical Ventilation
- Publication Year :
- 2015
-
Abstract
- Non-invasive ventilation performed through an oronasal mask is a standard in clinical and homecare mechanical ventilation. Besides all its advantages, inevitable leaks through the mask cause errors in the feedback information provided by the airflow sensor and, hence, patient-ventilator asynchrony with multiple negative consequences. Here we investigate a new way to provide a trigger to the ventilator. The method is based on the measurement of rib cage movement at the onset of inspiration and during breathing by fibre-optic sensors. In a series of simultaneous measurements by a long-period fibre grating sensor and pneumotachograph we provide the statistical evidence of the 200 ms lag of the pneumo with respect the fibre-optic signal. The lag is registered consistently across three independent delay metrics. Further, we discuss exceptions from this trend and identify the needed improvements to the proposed fibre-sensing scheme.