Back to Search Start Over

Preliminary design of a tangentially viewing imaging bolometer for NSTX-U

Authors :
Peterson, B. J.
Sano, R.
Reinke, M.L.
Canik, J.M.
Delgado-Aparicio, L. F.
Lore, J. D.
Mukai, K.
Gray, T.K.
van Eden, G.G.
Jaworski, M.A.
Peterson, B. J.
Sano, R.
Reinke, M.L.
Canik, J.M.
Delgado-Aparicio, L. F.
Lore, J. D.
Mukai, K.
Gray, T.K.
van Eden, G.G.
Jaworski, M.A.
Source :
Review of Scientific Instruments vol.87 (2016) nr.11 p.1-4 [ISSN 0034-6748]
Publication Year :
2016

Abstract

The infrared imaging video bolometer (IRVB) measures plasma radiated power images using a thin metal foil. Two different designs with a tangential view of NSTX-U are made assuming a 640 × 480 (1280 × 1024) pixel, 30 (105) fps, 50 (20) mK, IR camera imaging the 9 cm × 9 cm × 2 μm Pt foil. The foil is divided into 40 × 40 (64 × 64) IRVB channels. This gives a spatial resolution of 3.4 (2.2) cm on the machine mid-plane. The noise equivalent power density of the IRVB is given as 113 (46) μW/cm2 for a time resolution of 33 (20) ms. Synthetic images derived from Scrape Off Layer Plasma Simulation data using the IRVB geometry show peak signal levels ranging from ∼0.8 to ∼80 (∼0.36 to ∼26) mW/cm2.

Details

Database :
OAIster
Journal :
Review of Scientific Instruments vol.87 (2016) nr.11 p.1-4 [ISSN 0034-6748]
Notes :
Peterson, B. J.
Publication Type :
Electronic Resource
Accession number :
edsoai.on1019675377
Document Type :
Electronic Resource