Exploring novel sero-epidemiological tools—Effect of different storage conditions on longitudinal stability of microarray slides comprising influenza A-, measles- and Streptococcus pneumoniae antigens

Authors :
Freidl, G.S. (Gudrun)
Bruin, E. (Erwin) de
Schipper, M. (Maarten)
Koopmans D.V.M., M.P.G. (Marion)
Freidl, G.S. (Gudrun)
Bruin, E. (Erwin) de
Schipper, M. (Maarten)
Koopmans D.V.M., M.P.G. (Marion)
Publication Year :
2017

Abstract

In this study we evaluated the long-term stability of a microarray-based serological screening platform, containing antigens of influenza A, measles and Streptococcus pneumoniae, as part of a preparedness research program aiming to develop assays for syndromic disease detection. Spotted microarray slides were kept at four different storage regimes with varying temperature and humidity conditions. We showed that under the standard storage condition in a temperature-controlled (21 °C) and desiccated environment (0% relative humidity), microarray slides remained stable for at least 22 months without loss of antigen quality, whereas the other three conditions (37 °C, desiccated; Room temperature, non-desiccated; Frozen, desiccated) produced acceptable results for some antigens (influenza A, S.pneumoniae), but not for others (measles). We conclude that these arrays for multiplex antibody testing can be prepared and stored for prolonged periods of time, which aids laboratory-preparedness and facilitates sero-epidemiological studies.

Details

Database :
OAIster
Notes :
application/pdf, Journal of Virological Methods vol. 245, pp. 53-60, English
Publication Type :
Electronic Resource
Accession number :
edsoai.on1019674433
Document Type :
Electronic Resource
Full Text :
https://doi.org/10.1016.j.jviromet.2017.03.006