Back to Search Start Over

Development and Operation of Arrays of TES x-ray Microcalorimeters Suitable for Constellation-X

Authors :
Kilbourne, C. A
Bandler, S. R
Brown, A. D
Chervenak, J. A
Eckart, M. E
Finkbeiner, F. M
Iyomoto, N
Kelley, R. L
Porter, F. S
Smith, S. J
Irwin, K. D
Publication Year :
2008
Publisher :
United States: NASA Center for Aerospace Information (CASI), 2008.

Abstract

Having already developed a transition-edge-sensor (TES) microcalorimeter design that enables uniform and reproducible high spectral resolution (routinely better than 3 eV resolution at 6 keV) and is compatible with high fill-factor arrays, we are now working towards demonstrating this performance at high count rates and with the multiplexed read-out needed for instrumenting the Constellation-X X-ray Microcalorimeter Spectrometer (XMS) focal plane array. Design changes that increase the speed of the individual XMS pixels, such as lowering the heat capacity or increasing the thermal conductance of the link to the 50-mK heatsink, result in larger, faster signals, thus the coupling to the multiplexer and the overall bandwidth of the electronics must accommodate this increase in slew rate. In order to operate the array with high incident x-ray flux without unacceptable degradation of the spectral resolution, the magnitude of thermal and electrical crosstalk must be controlled. We will discuss recent progress in the thermal and electrical designs of our close-packed TES arrays, and we will present spectra acquired through the read-out chain from the multiplexer electronics, through the demultiplexer software, to real-time signal processing.

Details

Language :
English
Database :
NASA Technical Reports
Publication Type :
Report
Accession number :
edsnas.20080032783
Document Type :
Report