Dead-time effects in microchannel-plate imaging detectors
- Source :
- In: EUV, X-ray, and gamma-ray instrumentation for astronomy II; Proceedings of the Meeting, San Diego, CA, July 24-26, 1991 (A93-34376 13-35).
- Publication Year :
- 1991
- Publisher :
- United States: NASA Center for Aerospace Information (CASI), 1991.
-
Abstract
- The observed counting rates of microchannel plate (MCP) based detectors for high resolution observations of celestial EUV and X-ray sources vary over many orders of magnitude; the counting capability of an individual channel, however, is not high, and is associated with dead-times ranging from 0.1 msec to 1 sec. The dead-time increases with the area illuminated; attention is presently given to laboratory determinations of the count rate characteristics of a MCP detector as a function of illuminated area, and a model is developed for these results' use in the interpretation of space observations.