Back to Search Start Over

In situ ozone instrumentation for 10-Hz measurements - Development and evaluation

Authors :
Gregory, G. L
Hudgins, C. H
Ritter, J. A
Lawrence, M
Publication Year :
1987
Publisher :
United States: NASA Center for Aerospace Information (CASI), 1987.

Abstract

The development of the fast-response ozone detector for the Electra aircraft is described. The selection of a technique to meet the design goal of 10-Hz detection is examined in terms of detection principles, instrument sampling parameters, signal conditioning, and aircraft and sampling environment. An instrument which employs a NO technique for detection of ozone with a reaction chamber volume of 16 cu cm, a pressure of 60 torr, and a sample flow of 1000 standard cu cm/min was developed. Laboratory and flight testings of the detector were conducted in order to evaluate its performance. The data reveal that the fast-response ozone detector is highly reliable with a response of 0.1 sec to 90 percent of reading, has a lower detection limit of 1 ppbv, and an S/N of 20 at 20 ppbv ozone.

Subjects

Subjects :
Instrumentation And Photography

Details

Language :
English
Database :
NASA Technical Reports
Publication Type :
Report
Accession number :
edsnas.19870054911
Document Type :
Report