Back to Search
Start Over
Characterization of CdTe/CdZnTe detectors
- Source :
- IEEE Transactions on Nuclear Science. June, 2002, Vol. 49 Issue 3, p1258, 6 p.
- Publication Year :
- 2002
-
Abstract
- In order to characterize CdTe/CdZnTe detectors in a planar configuration, we have developed a new method to extract [mu][tau] products. In this method, we prepare an analytic spectral model based on the charge transport properties in the device, which is intended to be used in fitting calculation. The low mobility-lifetime ([mu][tau]) products of carriers in CdTe/CdZnTe detectors produce a position dependency in the charge induction efficiency. The model takes the induction efficiency and interaction positions of photons into account. Since the model is parameterized by [mu][tau] products, it can extract [mu][tau] products. Here, we demonstrate how the model works based on the results from 2-mm-thick HPB CdZnTe and THM CdTe detectors. Index Terms--CdTe, CdZnTe, gamma-ray detector, mobility-lifetime products, spectral modeling.
Details
- ISSN :
- 00189499
- Volume :
- 49
- Issue :
- 3
- Database :
- Gale General OneFile
- Journal :
- IEEE Transactions on Nuclear Science
- Publication Type :
- Academic Journal
- Accession number :
- edsgcl.93701778