Back to Search Start Over

A fast readout method for an imaging plate using wavelength shifting fibers and line-shaped laser

Authors :
Sakasai, K.
Katagiri, M.
Nakamura, T.
Toh, K.
Takahashi, H.
Nakazawa, M.
Source :
IEEE Transactions on Nuclear Science. Feb, 2002, Vol. 49 Issue 1, p156, 4 p.
Publication Year :
2002

Abstract

A fast readout method for an imaging plate is proposed using wavelength shifting fibers and line-shaped laser. This method adopts wavelength shifting fiber for signal transmission and line-shaped laser for reading out all pixels of the imaging plate in one axis at the same time, which results in reduction of readout time. To validate the method, the authors have fabricated an experimental system with a rotating mirror and carried out demonstration experiments using CAMAC readout system. In the experiments, two measuring method such as a photon counting method and a charge integral method were used. The experimental results show that an alpha particle image of 5 x 5 mm can be taken within 10 ms in both methods. The spatial resolution of 0.5 mm in the photon counting method was obtained in scanning direction, although the resolution perpendicular to scanning direction was more than 1 mm, which greatly depends on the diameter of the wavelength shifting fibers. Index Terms--Fast readout, imaging plate, line-shaped laser, wavelength shifting fibers.

Details

ISSN :
00189499
Volume :
49
Issue :
1
Database :
Gale General OneFile
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
edsgcl.86828493