Back to Search Start Over

Electrochemical Sensor Based on a Nanocomposite of Carbon Dots, Hexadecyltrimethylammonium Bromide and Chitosan for Mesalazine Determination

Authors :
Fahimeh Jalali
Hassanvand, Zahra
Barati, Ali
Source :
Journal of Analytical Chemistry. April, 2020, Vol. 75 Issue 4, p544, 9 p.
Publication Year :
2020

Abstract

A glassy carbon electrode modified with a nanocomposite consisting of carbon dots, hexadecyltrimethyl ammonium bromide and chitosan showed excellent electrocatalytic activity toward oxidation of mesalazine (MSA). Under optimal experimental conditions, hydrodynamic amperometry was used for sensitive determination of MSA. Calibration curve was obtained in the range of 0.1-10 [mu]M with a detection limit of 0.05 [mu]M. The repeatability of the method was studied for 5 replicate measurements of MSA (50 [mu]M) and the calculated relative standard deviation (RSD %) was 0.74%. The reproducibility for 5 different electrodes (expressed as RSD %) was 1.96% for MSA (50 [mu]M). The amperometric determination of the drug was successfully performed in spiked blood serum samples with recoveries ranging from 96 to 99%.<br />Author(s): Fahimeh Jalali [sup.1], Zahra Hassanvand [sup.1], Ali Barati [sup.1] Author Affiliations: (1) grid.412668.f, 0000 0000 9149 8553, Department of Chemistry, Razi University, , Kermanshah, Iran Pharmaceutical analysis is a [...]

Details

Language :
English
ISSN :
10619348
Volume :
75
Issue :
4
Database :
Gale General OneFile
Journal :
Journal of Analytical Chemistry
Publication Type :
Academic Journal
Accession number :
edsgcl.620954759
Full Text :
https://doi.org/10.1134/S1061934820040061