Back to Search Start Over

Measurements of X-ray imaging performance of granular phosphors with direct-coupled CMOS sensors

Authors :
Cho, Min Kook
Kim, Ho Kyung
Graeve, Thorsten
Yun, Seung Man
Lim, Chang Hwy
Cho, Hyosung
Kim, Jung-Min
Source :
IEEE Transactions on Nuclear Science. June, 2008, Vol. 55 Issue 3, p1338, 6 p.
Publication Year :
2008

Abstract

For [Gd.sub.2][O.sub.2]S:Tb granular phosphor screens having a wide range of mass thicknesses, we have investigated the fundamental imaging performance in terms of modulation-transfer function (MTF), noise-power spectrum (NPS) and detective quantum efficiency (DQE). As an optical photon readout device, a CMOS photodiode array with a pitch of 48 [micro]m was used. Under the representative radiation quality, RQA 5, recommended by the IEC (International Electrotechnical Commission, Report 1267), the MTF was measured using a slanted-slit method to avoid aliasing and the NPS was determined by two-dimensional (2D) Fourier analysis of white images. The DQE was assessed from the measured MTF, NPS and the estimated photon fluence. Figure-of-merit (FOM) curves are presented to describe the tradeoff between the X-ray sensitivity and spatial resolution of screens as a function of mass thickness. This study will be useful for the selection guidance of [Gd.sub.2][O.sub.2]S:Tb phosphors for the relevant imaging tasks of digital radiography. Index Terms--Detective quantum efficiency, digital radiography, granular phosphor, modulation-transfer function, noise-power spectrum.

Details

Language :
English
ISSN :
00189499
Volume :
55
Issue :
3
Database :
Gale General OneFile
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
edsgcl.180798615