Back to Search Start Over

Electrowetting-based microfluidics for analysis of peptides and proteins by matrix-assisted laser desorption/ionization mass spectrometry

Authors :
Wheeler, Aaron R.
Moon, Hyejin
Kim, Chang-Jin
Loo, Joseph A.
Garrell, Robin L.
Source :
Analytical Chemistry. August 15, 2004, Vol. 76 Issue 16, p4833, 6 p.
Publication Year :
2004

Abstract

A new technique for preparing samples for matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) is reported. The technique relies on electrowetting-on-dielectric (EWOD) to move droplets containing proteins or peptides and matrix to specific locations on an array of electrodes for analysis. Standard MALDI-MS reagents, analytes, concentrations, and recipes are demonstrated to be compatible with the technique. Mass spectra are comparable to those collected by conventional methods. Nonspecific adsorption of analytes to device surfaces is demonstrated to be negligible. The results suggest that EWOD may be a useful tool for automating sample preparation for high-throughput proteomics and other applications of MALDI-MS.