A new method for evaluation of cavitation near mechanical heart valves
- Source :
- Journal of Biomechanical Engineering. Oct, 2003, Vol. 125 Issue 5, p663, 8 p.
- Publication Year :
- 2003
-
Abstract
- Evaluation of cavitation in vivo is often based on recordings of high-pass filtered random high-frequency pressure fluctuations. We hypothesized that cavitation signal components are more appropriately assessed by a new method for extraction of random signal components of the pressure signals. We investigated three different valve types and found a high correlation between the two methods ([r.sup.2]:0.8806-0.9887). The new method showed that the cavitation signal could be extracted without a priori knowledge needed for setting the high-pass filter cut off frequency, nor did it introduce bandwidth limitation of the cavitation signal. [DOI: 10.1115/1.1613297]