Back to Search Start Over

Clinical utility of a rapid two-dimensional balanced steady-state free precession sequence with deep learning reconstruction

Authors :
Katerina Eyre
Moezedin Javad Rafiee
Margherita Leo
Junjie Ma
Elizabeth Hillier
Negin Amini
Josephine Pressacco
Martin A. Janich
Xucheng Zhu
Matthias G. Friedrich
Michael Chetrit
Source :
Journal of Cardiovascular Magnetic Resonance, Vol 26, Iss 2, Pp 101069- (2024)
Publication Year :
2024
Publisher :
Elsevier, 2024.

Abstract

ABSTRACT: Background: Cardiovascular magnetic resonance (CMR) cine imaging is still limited by long acquisition times. This study evaluated the clinical utility of an accelerated two-dimensional (2D) cine sequence with deep learning reconstruction (Sonic DL) to decrease acquisition time without compromising quantitative volumetry or image quality. Methods: A sub-study using 16 participants was performed using Sonic DL at two different acceleration factors (8× and 12×). Quantitative left-ventricular volumetry, function, and mass measurements were compared between the two acceleration factors against a standard cine method. Following this sub-study, 108 participants were prospectively recruited and imaged using a standard cine method and the Sonic DL method with the acceleration factor that more closely matched the reference method. Two experienced clinical readers rated images based on their diagnostic utility and performed all image contouring. Quantitative contrast difference and endocardial border sharpness were also assessed. Left- and right-ventricular volumetry, left-ventricular mass, and myocardial strain measurements were compared between cine methods using Bland-Altman plots, Pearson’s correlation, and paired t-tests. Comparative analysis of image quality was measured using Wilcoxon-signed-rank tests and visualized using bar graphs. Results: Sonic DL at an acceleration factor of 8 more closely matched the reference cine method. There were no significant differences found across left ventricular volumetry, function, or mass measurements. In contrast, an acceleration factor of 12 resulted in a 6% (5.51/90.16) reduction of measured ejection fraction when compared to the standard cine method and a 4% (4.32/88.98) reduction of measured ejection fraction when compared to Sonic DL at an acceleration factor of 8. Thus, Sonic DL at an acceleration factor of 8 was chosen for downstream analysis. In the larger cohort, this accelerated cine sequence was successfully performed in all participants and significantly reduced the acquisition time of cine images compared to the standard 2D method (reduction of 37% (5.98/16) p 0.05). Left- and right-ventricular volumetry and circumferential and radial strain were also similar between methods (p > 0.05) but left-ventricular mass and longitudinal strain were over-estimated using the proposed accelerated cine method (mass over-estimated by 3.36 g/m2, p

Details

Language :
English
ISSN :
10976647
Volume :
26
Issue :
2
Database :
Directory of Open Access Journals
Journal :
Journal of Cardiovascular Magnetic Resonance
Publication Type :
Academic Journal
Accession number :
edsdoj.fffdb7df29c443bdb7772845b16b8b43
Document Type :
article
Full Text :
https://doi.org/10.1016/j.jocmr.2024.101069