Accuracy and Reproducibility of a Modified Echocardiographic Method for Right Ventricular Output Calculation in Neonates
- Source :
- Journal of Cardiovascular Development and Disease, Vol 12, Iss 1, p 18 (2025)
- Publication Year :
- 2025
- Publisher :
- MDPI AG, 2025.
-
Abstract
- We aimed to evaluate the accuracy and reproducibility of right ventricular output (RVO) using different anatomical landmarks: the internal pulmonary valve diameter (PVD) between the valve hinge points (hinge-PVD) according to the traditional technique, and PVD between the valve leaflet tips (tip-PVD). This was a retrospective analysis of prospective collected data. All neonates with echocardiographic measurements of RVO and left ventricular output (LVO) without congenital heart disease, including patent ductus arteriosus and patent foramen ovale > 3 mm, were included. Accuracy was assessed by comparison with LVO. Intra- and inter-observer reproducibility of the off-line analysis were assessed. Forty-five neonates were included. RVO calculation with tip-PVD was more accurate than hinge-PVD in comparison with LVO, r2 0.712 versus 0.464, bias (95% limits of agreement) 1.4 mL/kg/min (−26–29 mL/kg/min) versus 61 mL/kg/min (−11–132 mL/kg/min), respectively. Both hinge-PVD and tip-PVD presented similar reproducibility, with an intra-observer bias (95% LOA) of 0.3 (−1.0–0.5) and –0.2 (−0.8–0.5) respectively, and an inter-observer bias of 0.1 (−1.3–1.6) and 0.1 (−1.4–1.6). RVO calculation using tip-PVD was more accurate than the conventional technique, with similar reproducibility.
Details
- Language :
- English
- ISSN :
- 23083425
- Volume :
- 12
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- Journal of Cardiovascular Development and Disease
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.f036660ca9e4433390e726470a8eb148
- Document Type :
- article
- Full Text :
- https://doi.org/10.3390/jcdd12010018