Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study

Authors :
Djamel Dabli
Cécile Salvat
Isabelle Fitton
Claire Van Ngoc Ty
Peggy Palanchon
Jean-Paul Beregi
Joël Greffier
Lama Hadid-Beurrier
Source :
Sensors, Vol 24, Iss 21, p 6883 (2024)
Publication Year :
2024
Publisher :
MDPI AG, 2024.

Abstract

An image-quality CT phantom was scanned with three different 3D X-ray imaging guidance devices in the operating theatre: O-Arm, Loop-X, and Airo TruCT. Default acquisition and reconstruction parameters for lumbar spine procedures were used on each device. The tube current was set to a dose level of around 27 mGy. A task-based image quality assessment was performed by calculating the noise power spectrum (NPS) and task transfer function (TTF). A detectability index (d’) was calculated for three simulated bone lesions. The noise magnitude of the O-Arm was higher than the Airo TruCT, and the Loop-X had higher noise than the Airo TruCT. The highest average NPS frequency was for the O-Arm images, and the lowest was for the Loop-X. The TTFs at 50% values were similar for the Airo TruCT and Loop-X devices. Compared to Airo TruCT, the TTF at 50% value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. Compared to Airo TruCT, the d’ value was lower with Loop-X by −26.73%, −27.02%, and −23.95% for lytic lesions, sclerotic lesions, and high-density bone, respectively. Each 3D-imaging spine surgery guidance device has its own strengths and weaknesses in terms of image quality. Cone-beam CT systems apparently offer the best compromise between noise and spatial resolution for spine surgery.

Details

Language :
English
ISSN :
14248220
Volume :
24
Issue :
21
Database :
Directory of Open Access Journals
Journal :
Sensors
Publication Type :
Academic Journal
Accession number :
edsdoj.bfc48f60edba4da4a6453fea13845eed
Document Type :
article
Full Text :
https://doi.org/10.3390/s24216883