Back to Search
Start Over
Comparison of automated corneal endothelial cell analysis in healthy and postoperative eyes with phakic intraocular lens: a cross-sectional study and literature review
- Source :
- BMC Ophthalmology, Vol 24, Iss 1, Pp 1-11 (2024)
- Publication Year :
- 2024
- Publisher :
- BMC, 2024.
-
Abstract
- Abstract Purpose Measurement of corneal endothelial cells is critical for postoperative evaluation of phakic intraocular lens (pIOL) surgery. However, inter-instrument differences in corneal endothelial cell density (ECD) after pIOL implantation have not yet been reported. This study aimed to compare automated corneal endothelial cell analysis between CellChek-20 (Konan Medical, Hyogo, Japan) and EM-4000 (Tomey, Nagoya, Japan) in healthy and postoperative eyes with pIOL. Methods We retrospectively analyzed 154 healthy and 236 postoperative eyes after pIOL surgery. Endothelial cell measurements were performed using CellChek-20 and EM-4000 with autofocusing and automated image analysis. ECD, percentage of hexagonal cells (HEX), coefficient of variation in cell size (CoV), and central corneal thickness (CCT) were compared between the two devices. Results The ECDs of the two devices were highly correlated in both healthy (Spearman’s correlation coefficient [r] = 0.805; p
Details
- Language :
- English
- ISSN :
- 14712415
- Volume :
- 24
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- BMC Ophthalmology
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.bccd13495a140d7be811feaadd921f7
- Document Type :
- article
- Full Text :