Back to Search Start Over

Studies on the Aptasensor Miniaturization for Electrochemical Detection of Lead Ions

Authors :
Marta Jarczewska
Marta Sokal
Marcin Olszewski
Elzbieta Malinowska
Source :
Biosensors, Vol 14, Iss 2, p 110 (2024)
Publication Year :
2024
Publisher :
MDPI AG, 2024.

Abstract

Lead poses severe effects on living organisms, and since Pb2+ ions tend to accumulate in different organs, it is crucial to monitor Pb2+ concentration in samples such as water and soil. One of the approaches is the utilization of biosensors combined with aptamer-based layers for the electrochemical detection of lead ions. Herein, we present the studies of applying miniaturized screen-printed transducers as solid surfaces to fabricate aptamer layers. As the research is the direct continuation of our previous studies regarding the use of gold disk electrodes, the working parameters of elaborated aptasensors were defined, including the range of linear response (10–100 nM), selectivity as well as stability, regeneration, and feasibility of application for the analysis of real samples. This was achieved using voltammetric techniques including cyclic and square-wave voltammetry in the presence of methylene blue redox indicator.

Details

Language :
English
ISSN :
20796374
Volume :
14
Issue :
2
Database :
Directory of Open Access Journals
Journal :
Biosensors
Publication Type :
Academic Journal
Accession number :
edsdoj.ba256fae2b424568bd99e489953104c6
Document Type :
article
Full Text :
https://doi.org/10.3390/bios14020110