Back to Search Start Over

Estudo biomecânico da fixação transcortical ou transesponjosa do enxerto de tendão patelar com pinos bioabsorvíveis na reconstrução do LCA em ovinos Biomechanical study of transcortical or transtrabecular bone fixation of patellar tendon graft with bioabsorbable pins in ACL reconstruction in sheep

Authors :
Mauro Batista Albano
Paulo César Borges
Mario Massatomo Namba
João Luiz Vieira da Silva
Francisco de Assis Pereira Filho
Edmar Stieven Filho
Jorge Eduardo Fouto Matias
Source :
Revista Brasileira de Ortopedia, Vol 47, Iss 1, Pp 43-49 (2012)
Publication Year :
2012
Publisher :
Thieme Revinter Publicações Ltda., 2012.

Abstract

OBJETIVO: Determinar a resistência inicial de fixação do sistema RigidFix® e comparar com o método tradicional de fixação que utiliza parafusos de interferência metálicos. Avaliar a resistência da fixação com o sistema RigidFix® ao se modificar o posicionamento rotacional do bloco ósseo no interior do túnel femoral. MÉTODOS: 40 espécimes de joelhos (soldras) de ovinos, foram submetidos à reconstrução do ligamento cruzado anterior (LCA)utilizando enxerto osso-tendão-osso. Em 20 espécimes utilizou-se o método RigidFix®, este grupo foi subdividido em dois, 10 joelhos foram utilizados para fixação através da esponjosa e 10 para fixação cruzando a tábua óssea cortical. Nos 20 espécimes restantes fixou-se o enxerto com parafusos de interferência metálico de 9mm. RESULTADOS: A comparação do método RigidFix® com o método de fixação com parafuso de interferência metálico não mostrou diferenças estatisticamente significativas ao se considerar carga máxima e rigidez, também não houve diferenças estatisticamente significativas ao se modificar o posicionamento rotacional do bloco ósseo no interior do túnel femoral. Para estas avaliações considerou-se o p < 0,017. CONCLUSÃO: A fixação do enxerto osso-tendão-osso com dois pinos bioabsorvíveis, independente do posicionamento rotacional no túnel femoral, permite uma fixação comparável em termos de resistência inicial ao parafuso de interferência metálico neste modelo experimental.OBJECTIVE: to determine the initial resistance of fixation using the Rigid Fix® system, and compare it with traditional fixation methods using metal interference screws; and to evaluate the resistance of the fixation with the rigid fix system when the rotational position of the bone block is altered in the interior of the femoral tunnel. METHODS: forty ovine knee specimens (stifle joints) were submitted to anterior cruciate ligament reconstruction (ACL) using a bone-tendon-bone graft. In twenty specimens, the RigidFix® method was used; this group was subdivided into two groups: ten knees used for fixation with trabecular bone screw, and ten for fixation passing through the layer of cortical bone. In the twenty remaining specimens, the graft was fixed with 9mm metal interference screws. RESULTS: comparison of the Rigidfix® method with the metal interference screw fixation method did not show any statistically significant differences in terms of maximum load and rigidity; also, there were no statistically significant differences when the rotational position of the bone block was altered inside the femoral tunnel. For these evaluations, a level of significance of p < 0.017 was considered. CONCLUSION: fixation of the bone-tendon-bone graft with 2 bioabsorbable pines, regardless of the rotational position inside the femoral tunnel, gave a comparable fixation in terms of initial resistance to the metal interference screw, in this experimental model.

Details

Language :
English, Portuguese
ISSN :
01023616 and 19824378
Volume :
47
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Revista Brasileira de Ortopedia
Publication Type :
Academic Journal
Accession number :
edsdoj.b724d3c1169e46f2b2e50386cea5ccfb
Document Type :
article
Full Text :
https://doi.org/10.1590/S0102-36162012000100006