Experimental application of an automated alignment correction algorithm for geological CT imaging: phantom study and application to sediment cores from cold-water coral mounds

Authors :
Stephan Skornitzke
Jacek Raddatz
André Bahr
Gregor Pahn
Hans-Ulrich Kauczor
Wolfram Stiller
Source :
European Radiology Experimental, Vol 3, Iss 1, Pp 1-8 (2019)
Publication Year :
2019
Publisher :
SpringerOpen, 2019.

Abstract

Abstract Background In computed tomography (CT) quality assurance, alignment of image quality phantoms is crucial for quantitative and reproducible evaluation and may be improved by alignment correction. Our goal was to develop an alignment correction algorithm to facilitate geological sampling of sediment cores taken from a cold-water coral mount. Methods An alignment correction algorithm was developed and tested with a CT acquisition at 120 kVp and 150 mAs of an image quality phantom. Random translation (maximum 15 mm) and rotation (maximum 2.86°) were applied and ground-truth was compared to parameters determined by alignment correction. Furthermore, mean densities were evaluated in four regions of interest (ROIs) placed in the phantom low-contrast section, comparing values before and after correction to ground truth. This process was repeated 1000 times. After validation, alignment correction was applied to CT acquisitions (140 kVp, 570 mAs) of sediment core sections up to 1 m in length, and sagittal reconstructions were calculated for sampling planning. Results In the phantom, average absolute differences between applied and detected parameters after alignment correction were 0.01 ± 0.06 mm (mean ± standard deviation) along the x-axis, 0.11 ± 0.08 mm along the y-axis, 0.15 ± 0.07° around the x-axis, and 0.02 ± 0.02° around the y-axis, respectively. For ROI analysis, differences in densities were 63.12 ± 30.57, 31.38 ± 32.10, 18.27 ± 35.57, and 9.59 ± 26.37 HU before alignment correction and 1.22 ± 1.40, 0.76 ± 0.9, 0.45 ± 0.86, and 0.36 ± 0.48 HU after alignment correction, respectively. For sediment core segments, average absolute detected parameters were 3.93 ± 2.89 mm, 7.21 ± 2.37 mm, 0.37 ± 0.33°, and 0.21 ± 0.22°, respectively. Conclusions The alignment correction algorithm was successfully evaluated in the phantom and allowed a correct alignment of sediment core segments, thus aiding in sampling planning. Application to other tasks, like image quality analysis, seems possible.

Details

Language :
English
ISSN :
25099280
Volume :
3
Issue :
1
Database :
Directory of Open Access Journals
Journal :
European Radiology Experimental
Publication Type :
Academic Journal
Accession number :
edsdoj.b2de046170e140c6a915451cbb54db75
Document Type :
article
Full Text :
https://doi.org/10.1186/s41747-019-0091-8