Back to Search Start Over

Factors associated with damages and repair costs of reusable flexible bronchoscopes in pediatrics in China

Authors :
Ruixue Hu
Yanhua Chen
Sixin Jiang
Ting Hu
Juanli Huang
Lu Zhang
Wei Pan
Liangying Yi
Yan Huang
Source :
BMC Pediatrics, Vol 24, Iss 1, Pp 1-9 (2024)
Publication Year :
2024
Publisher :
BMC, 2024.

Abstract

Abstract Background Frequent repairs of pediatric flexible bronchoscopes can lead to a huge financial burden for the hospital. This study aimed to investigate the common causes of the failures in pediatric flexible bronchoscopes and propose the measures to prevent the failures. Methods This was a retrospective study. We collected repair information of the pediatric flexible bronchoscopes reprocessed in the Department of Sterile Processing at a hospital between September 1, 2018 and September 1, 2022 in order to investigate the causes and possible factors associated with the failures in pediatric flexible bronchoscopes. Results The Department of Sterile Processing staff reprocessed the pediatric flexible bronchoscopes 4280 times. A total of 29 failures were identified. The failure rate was 0.678%. The average repair cost was USD7246.60. The common failures in the pediatric flexible bronchoscopes included dim video image, black dots, improper video image display or no image during angulation adjustment, and pressure marks in the insertion tube. The failure rates in flexible electronic bronchoscopes and small-diameter flexible bronchoscopes were 65.5% and 93.1%, respectively. The failure rate in the pediatric flexible bronchoscopes reprocessed by the staff members with less work experience was 75.9%. Conclusion The failure rate in the pediatric flexible bronchoscopes was not high but the repair costs were extremely high. The types and size of the flexible bronchoscopes and work experience of the staff members responsible for bronchoscope reprocessing were the possible factors associated with the failure rate in the pediatric flexible bronchoscopes. It is advisable to further optimize the central workflow and management mode for reprocessing the pediatric flexible bronchoscopes, thereby extending their useful life and reducing costs.

Details

Language :
English
ISSN :
14712431
Volume :
24
Issue :
1
Database :
Directory of Open Access Journals
Journal :
BMC Pediatrics
Publication Type :
Academic Journal
Accession number :
edsdoj.9e60e523366e4724ae9b719507441519
Document Type :
article
Full Text :
https://doi.org/10.1186/s12887-024-04705-2