Defibrillation effectiveness and safety of the shock waveform used in a contemporary wearable cardioverter defibrillator: Results from animal and human studies.
- Source :
- PLoS ONE, Vol 18, Iss 3, p e0281340 (2023)
- Publication Year :
- 2023
- Publisher :
- Public Library of Science (PLoS), 2023.
-
Abstract
- IntroductionThe wearable cardioverter defibrillator (WCD) is used to protect patients at risk for sudden cardiac arrest. We examined defibrillation efficacy and safety of a biphasic truncated exponential waveform designed for use in a contemporary WCD in three animal studies and a human study.MethodsAnimal (swine) studies: #1: Efficacy comparison of a 170J BTE waveform (SHOCK A) to a 150J BTE waveform (SHOCK B) that approximates another commercially available waveform. Primary endpoint first shock success rate. #2: Efficacy comparison of the two waveforms at attenuated charge voltages in swine at three prespecified impedances. Primary endpoint first shock success rate. #3: Safety comparison of SHOCK A and SHOCK B in swine. Primary endpoint cardiac biomarker level changes baseline to 6 and 24 hours post-shock. Human Study: Efficacy comparison of SHOCK A to prespecified goal and safety evaluation. Primary endpoint cumulative first and second shock success rate. Safety endpoint adverse events.ResultsAnimal Studies #1: 120 VF episodes in six swine. First shock success rates for SHOCK A and SHOCK B were 100%; SHOCK A non-inferior to SHOCK B (entire 95% CI of rate difference above -10% margin, p < .001). #2: 2,160 VF episodes in thirty-six swine. Attenuated SHOCK A was non-inferior to attenuated SHOCK B at each impedance (entire 95% CI of rate difference above -10% margin, p < .001). #3: Ten swine, five shocked five times each with SHOCK A, five shocked five times each with SHOCK B. No significant difference in troponin I (p = 0.658) or creatine phosphokinase (p = 0.855) changes from baseline between SHOCK A and SHOCK B. Human Study: Thirteen patients, 100% VF conversion rate. Mild skin irritation from adhesive defibrillation pads in three patients.ConclusionsThe BTE waveform effectively and safely terminated induced VF in swine and a small sample in humans.Trial registrationHuman study clinical trial registration: URL: https://clinicaltrials.gov; Unique identifier: NCT04132466.
Details
- Language :
- English
- ISSN :
- 19326203
- Volume :
- 18
- Issue :
- 3
- Database :
- Directory of Open Access Journals
- Journal :
- PLoS ONE
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.9c332b0a79f459ebeaf72b189a5c3df
- Document Type :
- article
- Full Text :
- https://doi.org/10.1371/journal.pone.0281340