Back to Search
Start Over
Comparing automated and manual assessments of tear break-up time using different non-invasive devices and a fluorescein procedure
- Source :
- Scientific Reports, Vol 14, Iss 1, Pp 1-12 (2024)
- Publication Year :
- 2024
- Publisher :
- Nature Portfolio, 2024.
-
Abstract
- Abstract To evaluate the agreement and repeatability of an automated topography-based method for non-invasive break-up time (NIBUT) analyses in comparison with two other NIBUT procedures, the fluorescein procedure (fBUT), and with the manual assessment with the same device. In the first experiment, a semi-randomised crossover study was performed on forty-three participants (23.1 ± 2.1 years). NIBUT measurements were collected in a randomised order, in both eyes of participants with EasyTear View + (Easytear, Rovereto), Polaris, and Sirius + (CSO, Firenze). Then a fBUT was collected. The overall measurement procedure was repeated in a further session (retest) on the same day. In a second experiment, a retrospective randomised crossover study was performed on eighty-five NIBUT videos previously recorded by the Sirius+. Two observers assessed manually the videos and the NIBUTs were compared with the automatic ones. In the first experiment, ANOVA showed a significant difference between the four measures in both eyes (p
Details
- Language :
- English
- ISSN :
- 20452322
- Volume :
- 14
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- Scientific Reports
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.9a7ee9fd5224e26932ca808e07aefe1
- Document Type :
- article
- Full Text :
- https://doi.org/10.1038/s41598-024-52686-0