Role of home monitoring with iCare ONE rebound tonometer in glaucoma patients management
- Source :
- International Journal of Ophthalmology, Vol 14, Iss 3, Pp 405-408 (2021)
- Publication Year :
- 2021
- Publisher :
- Press of International Journal of Ophthalmology (IJO PRESS), 2021.
-
Abstract
- AIM: To evaluate intraocular pressure (IOP) measurements and fluctuations using the iCare ONE rebound tonometer (RT-ONE), during home monitoring, in diagnosed and suspected glaucoma patients. METHODS: A retrospective case series of consecutive patients with known glaucoma or glaucoma suspects who were followed-up and treated between January 2016 and January 2017. The study included 80 eyes of 40 patients with a mean age of 59.1±14.6y (range, 24-78). All patients have undergone 4-5d of IOP home monitoring with RT-ONE at morning, noon, afternoon, and night time. RESULTS: Baseline mean IOP, as measured in the clinic (8 a.m.-12 p.m.), was 17.4±5.1 mm Hg, compared to RT-ONE home monitoring mean IOP of 15.6±4.1 mm Hg (P=0.002). Mean IOP was significantly lower at noon, afternoon and night times compared to clinic measured IOP and morning measurements (P=0.005). IOP peak measured during home monitoring was significantly higher compared to the clinic measured IOP (21.3±5.6 mm Hg and 17.4±5.1 mm Hg, P
- Subjects :
- glaucoma
intraocular pressure
home monitoring
icare one
peak
Ophthalmology
RE1-994
Subjects
Details
- Language :
- English
- ISSN :
- 22223959 and 22274898
- Volume :
- 14
- Issue :
- 3
- Database :
- Directory of Open Access Journals
- Journal :
- International Journal of Ophthalmology
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.924e5a81ece84d4394668c6ee275670d
- Document Type :
- article
- Full Text :
- https://doi.org/10.18240/ijo.2021.03.12