X-ray Spectrum Reconstruction by Diamond Detectors with Linear Response to Dose Rate

Authors :
Daniele Maria Trucchi
Paolo Ascarelli
Source :
Crystals, Vol 11, Iss 10, p 1258 (2021)
Publication Year :
2021
Publisher :
MDPI AG, 2021.

Abstract

The absorbers method is here applied by interposing filters of variable thickness between the X-ray source and a detector so to attenuate the radiation intensity by using the attenuation coefficient as a selective photon energy operator. The analysis of the signal provided by a polycrystalline diamond thin film detector exposed to the energy-selectively-attenuated X-ray beam was used for the reconstruction of the radiation spectrum. The 50 μm thick diamond detector achieves conditions of linear response to the dose rate of the incident radiation (linearity coefficient of 0.997 ± 0.003) for a bias voltage ≥90 V, corresponding to an electric field ≥1.8 × 104 V/cm. Once the absorbers method is applied, only the detector signal linearity to dose rate allows reconstructing the source X-ray bremsstrahlung spectrum with sufficiently high accuracy.