Back to Search Start Over

Turbolysis: A low-cost, small footprint alternative to commercial bead beaters for cell lysis

Authors :
Jason D. Limberis
John Z. Metcalfe
Source :
HardwareX, Vol 19, Iss , Pp e00576- (2024)
Publication Year :
2024
Publisher :
Elsevier, 2024.

Abstract

turboLysis is a novel mechanical cell lysis device that utilizes small beads to efficiently lyse tough cells like Mycobacterium, Saccharomyces, and Arabidopsis. We compared turboLysis to bead beating using the BeadBug 6 for several concentrations of Mycobacterium tuberculosis roughly correlated to the bacterial load commonly seen in patient samples. turboLysis performed similarly to the BeadBug at low bacterial concentrations and outperformed it at high concentrations above 2x105 CFU/ml (p

Details

Language :
English
ISSN :
24680672
Volume :
19
Issue :
e00576-
Database :
Directory of Open Access Journals
Journal :
HardwareX
Publication Type :
Academic Journal
Accession number :
edsdoj.8eb386147ac746f78073ea43b29db2e0
Document Type :
article
Full Text :
https://doi.org/10.1016/j.ohx.2024.e00576