Back to Search
Start Over
Beam output checks of a commercial high-field magnetic resonance-guided radiotherapy machine with its on-board megavoltage imager
- Source :
- Physics and Imaging in Radiation Oncology, Vol 25, Iss , Pp 100411- (2023)
- Publication Year :
- 2023
- Publisher :
- Elsevier, 2023.
-
Abstract
- Beam output checks of a commercial high-field magnetic resonance-guided radiotherapy machine can be performed with its on-board megavoltage imager (MVI). This is a fast and efficient method, but only recommended for daily checks. The aim of our study was to show its suitability for weekly checks by investigating its long-term agreement with the golden standard: ionization chamber measurements in a water tank. For one year, the output deviations obtained with both methods were compared. The difference was 0.1 ± 0.3 (1SD) percentage point. This indicated an excellent agreement, and translated into a tolerance level of ± 2 %.
Details
- Language :
- English
- ISSN :
- 24056316
- Volume :
- 25
- Issue :
- 100411-
- Database :
- Directory of Open Access Journals
- Journal :
- Physics and Imaging in Radiation Oncology
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.82a48ee3a65d4489a4f201766f5aceac
- Document Type :
- article
- Full Text :
- https://doi.org/10.1016/j.phro.2023.100411