Back to Search Start Over

Proton Beam Range and Charge Verification Using Multilayer Faraday Collector

Authors :
Ping L Yeap MPhil
Kah S Lew BSc
Wei Y C Koh PhD
Clifford G A Chua BSc
Andrew Wibawa MSc
Zubin Master MSc
James C L Lee PhD
Sung Y Park PhD
Hong Q Tan PhD
Source :
Technology in Cancer Research & Treatment, Vol 23 (2024)
Publication Year :
2024
Publisher :
SAGE Publishing, 2024.

Abstract

Purpose A daily quality assurance (QA) check in proton therapy is ensuring that the range of each proton beam energy in water is accurate to 1 mm. This is important for ensuring that the tumor is adequately irradiated while minimizing damage to surrounding healthy tissue. It is also important to verify the total charge collected against the beam model. This work proposes a time-efficient method for verifying the range and total charge of proton beams at different energies using a multilayer Faraday collector (MLFC). Methods We used an MLFC-128-250 MeV comprising 128 layers of thin copper foils separated by thin insulating Kapton TM layers. Protons passing through the collector induce a charge on the metallic foils, which is integrated and measured by a multichannel electrometer. The charge deposition on the foils provides information about the beam range. Results Our results show that the proton beam range obtained using MLFC correlates closely with the range obtained from commissioning water tank measurements for all proton energies. Upon applying a range calibration factor, the maximum deviation is 0.4 g/cm 2 . The MLFC range showed no dependence on the number of monitor units and the source-to-surface distance. Range measurements collected over multiple weeks exhibited stability. The total charge collected agrees closely with the theoretical charge from the treatment planning system beam model for low- and mid-range energies. Conclusions We have calibrated and commissioned the use of the MLFC to easily verify range and total charge of proton beams. This tool will improve the workflow efficiency of the proton QA.

Details

Language :
English
ISSN :
15330338
Volume :
23
Database :
Directory of Open Access Journals
Journal :
Technology in Cancer Research & Treatment
Publication Type :
Academic Journal
Accession number :
edsdoj.7dd3dc9f347c4b82a4bfecdd9dd6c37c
Document Type :
article
Full Text :
https://doi.org/10.1177/15330338241262610