Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas
- Source :
- Therapeutic Advances in Ophthalmology, Vol 12 (2020)
- Publication Year :
- 2020
- Publisher :
- SAGE Publishing, 2020.
-
Abstract
- Purpose: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. Methods: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. Results: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p
- Subjects :
- Ophthalmology
RE1-994
Subjects
Details
- Language :
- English
- ISSN :
- 25158414
- Volume :
- 12
- Database :
- Directory of Open Access Journals
- Journal :
- Therapeutic Advances in Ophthalmology
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.7cfaefbc3a4844cb91985ea85eb1ce87
- Document Type :
- article
- Full Text :
- https://doi.org/10.1177/2515841420923190