Back to Search Start Over

Feasibility of low-dose CT with model-based iterative image reconstruction in follow-up of patients with testicular cancer

Authors :
Kevin P. Murphy
Lee Crush
Siobhan B. O’Neill
James Foody
Micheál Breen
Adrian Brady
Paul J. Kelly
Derek G. Power
Paul Sweeney
Jackie Bye
Owen J. O’Connor
Michael M. Maher
Kevin N. O’Regan
Source :
European Journal of Radiology Open, Vol 3, Iss , Pp 38-45 (2016)
Publication Year :
2016
Publisher :
Elsevier, 2016.

Abstract

Purpose: We examine the performance of pure model-based iterative reconstruction with reduced-dose CT in follow-up of patients with early-stage testicular cancer. Methods: Sixteen patients (mean age 35.6 ± 7.4 years) with stage I or II testicular cancer underwent conventional dose (CD) and low-dose (LD) CT acquisition during CT surveillance. LD data was reconstructed with model-based iterative reconstruction (LD–MBIR). Datasets were objectively and subjectively analysed at 8 anatomical levels. Two blinded clinical reads were compared to gold-standard assessment for diagnostic accuracy. Results: Mean radiation dose reduction of 67.1% was recorded. Mean dose measurements for LD–MBIR were: thorax – 66 ± 11 mGy cm (DLP), 1.0 ± 0.2 mSv (ED), 2.0 ± 0.4 mGy (SSDE); abdominopelvic – 128 ± 38 mGy cm (DLP), 1.9 ± 0.6 mSv (ED), 3.0 ± 0.6 mGy (SSDE). Objective noise and signal-to-noise ratio values were comparable between the CD and LD–MBIR images. LD–MBIR images were superior (p

Details

Language :
English
ISSN :
23520477
Volume :
3
Issue :
38-45
Database :
Directory of Open Access Journals
Journal :
European Journal of Radiology Open
Publication Type :
Academic Journal
Accession number :
edsdoj.7c84d692976a42df94e39cfa30f733fe
Document Type :
article
Full Text :
https://doi.org/10.1016/j.ejro.2016.01.002