Back to Search Start Over

Simultaneous Double Dose Measurements Using TLD-100H

Authors :
Giuseppe Stella
Antouman Sallah
Rosaria Galvagno
Alessia D’Anna
Anna Maria Gueli
Source :
Crystals, Vol 14, Iss 7, p 603 (2024)
Publication Year :
2024
Publisher :
MDPI AG, 2024.

Abstract

Thermoluminescent dosimeters (TLD) and optically stimulated luminescent dosimeters (OSLD) are practical, accurate, and precise tools for point dosimetry in medical physics applications. The objective of this study is to investigate the luminescence properties—both OSL and TL—of lithium fluoride (LiF) doped with magnesium (Mg), copper (Cu), and phosphorous (P) (LiF: Mg, Cu, P), commercially known as TLD-100H. The goal is to devise a methodological approach for dose measurement that allows for obtaining two independently measured dose values at each irradiation point, thereby improving accuracy and precision. The luminescence properties of TLD-100H were studied using a beta irradiation source (90Sr/90Y) integrated into the TL/OSL DA-15 automated Risø reader. This study identified the ideal experimental conditions for optimal dose evaluation and used them for dosimeter calibration across doses ranging from 0.5 to 4.0 Gy. The results demonstrated that, under optimal measurement parameters, the OSL and residual thermoluminescence (ResTL) signals—correlated to two trap systems within the dosimeter—exhibited high reproducibility, stability over multiple cycles, and high precision and accuracy (≤2%). Specifically, the OSL response showed good linear behavior across the investigated dose range, while the ResTL signal exhibited linear behavior between 0.5 and 2 Gy and sublinear behavior for doses >2 Gy.

Details

Language :
English
ISSN :
20734352
Volume :
14
Issue :
7
Database :
Directory of Open Access Journals
Journal :
Crystals
Publication Type :
Academic Journal
Accession number :
edsdoj.7af206fdffcc46fcb022620ab6a10ec8
Document Type :
article
Full Text :
https://doi.org/10.3390/cryst14070603