Back to Search Start Over

Pullout strength of different pedicle screws after primary and revision insertion: an in vitro study on polyurethane foam

Authors :
Lien-Chen Wu
Yueh-Ying Hsieh
Fon-Yih Tsuang
Yi-Jie Kuo
Chia-Hsien Chen
Chang-Jung Chiang
Source :
BMC Musculoskeletal Disorders, Vol 24, Iss 1, Pp 1-8 (2023)
Publication Year :
2023
Publisher :
BMC, 2023.

Abstract

Abstract Background Surgeons are routinely required to remove loose or failed pedicle screws and insert a new screw in their place. However, inserting a new screw into an existing hole may compromise the holding capacity of the pedicle screw. The purpose of this study is to evaluate the pullout strength of pedicle screws with different thread designs after the primary insertion and revision surgery in a synthetic bone model. Methods Four pedicle screws with different thread designs (single-lead-thread (SLT) screw, dual-lead-thread (DLT) screw, mixed-single-lead-thread (MSLT) screw, and proximal-unthreaded-dual-thread (PUDL) screw) were inserted into pre-drilled, untapped holes (ø 4.2 mm, length 35 mm) in Sawbone blocks of density 20 pcf. In the first sequence, a 6.0 mm screw was inserted into the predrilled foam block and the primary pullout strength of the screw was measured according to ASTM F543. In the second sequence, a 6.0 mm screw was inserted and removed, and then either a 6.5 mm screw of the same design or a different screw design was inserted into the same hole and the pullout strength recorded. Results In the first sequence, the mean pullout strength of the MSLT screw was significantly (p

Details

Language :
English
ISSN :
14712474
Volume :
24
Issue :
1
Database :
Directory of Open Access Journals
Journal :
BMC Musculoskeletal Disorders
Publication Type :
Academic Journal
Accession number :
edsdoj.78934e6825fe4f3993c8af8f1ed69e9d
Document Type :
article
Full Text :
https://doi.org/10.1186/s12891-023-07015-3