Evaluation of thermal effects of surgical energy devices: ex vivo study

Authors :
Toshiharu Morikawa
Shuzo Hamamoto
Masakazu Gonda
Kazumi Taguchi
Rei Unno
Koei Torii
Masahiko Isogai
Kengo Kawase
Takashi Nagai
Shoichiro Iwatsuki
Toshiki Etani
Taku Naiki
Atsushi Okada
Takahiro Yasui
Source :
Scientific Reports, Vol 14, Iss 1, Pp 1-9 (2024)
Publication Year :
2024
Publisher :
Nature Portfolio, 2024.

Abstract

Abstract This study evaluated the direct and indirect thermal effects of various surgical energy devices using an ex-vivo model. Two types of three devices were evaluated: ENSEAL™ X1 Curved Jaw Tissue Sealer (X1) and ENSEAL™ G2 Curved Tissue Sealer (G2) as vessel sealing systems (VSSs), and HARMONIC® HD1000i Shears (HA) as an ultrasonic activating device (USAD). Each device was activated once under DRY or WET conditions. The tissue’s maximum temperature (MT), steam MT surrounding the activation site, and steam spread area (SSA) were measured. Under WET conditions, the median MT of a porcine common carotid artery at 1 mm from the activation site by X1, G2, and HA were 84.4, 83.3, and 50.5 °C, respectively. The direct thermal effect of HA was the lowest among the three devices. VSSs showed higher tissue MT under WET conditions compared with DRY conditions. Conversely, USAD showed the opposite trend. G2 demonstrated a significantly higher MT than X1 and HA (P

Details

Language :
English
ISSN :
20452322
Volume :
14
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
edsdoj.6cdd5a3a281a4fe19036fdda9a9e45dc
Document Type :
article
Full Text :
https://doi.org/10.1038/s41598-024-78624-8