Analysis of Commissioning Parameters and its Validation of O-Ring Gantry Based Medical Linear Accelerator HalcyonTM for Improved Radiotherapy Technique

Authors :
ARVIND KUMAR
KIRAN SHARMA
CHANDI BHATT
Suresh Yadav
SWATI VERMA
SOUGATA BASU
POOJA DHAUNDIYAL
DEEPAK KAUSHIK
DINESH SAROJ
Manoj Gupta
Source :
Iranian Journal of Medical Physics, Vol 21, Iss 1, Pp 30-39 (2024)
Publication Year :
2024
Publisher :
Mashhad University of Medical Sciences, 2024.

Abstract

Introduction: Popularly, teletherapy (telecobalt/LA) equipment is based on a C-arm gantry system. Recently, a fast O-ring gantry system introduced a medical linear accelerator (LA) to smoothen the workflow of treatment of cancer patients because of the increasing trend of the number of cancer cases over the past few years. This study aimed to analyze the commissioning parameters and validation of the O-ring gantry-based LA for improved radiotherapy techniques. Material and Methods: Three-dimensional (3D) radiation field analyzer (RFA) used to commission HalcyonTM LA. It is used for measuring percent depth dose (PDD), profiles, and output factors. Results: TPS data was validated by comparing it with our measured data. Plans per the TG-119 protocol showed good agreement between treatment planning systems (TPS) calculated and measured doses. For patient-specific, QA plans showed good agreement with gamma evaluation criteria. Conclusion: Commissioning and validation of O-ring gantry system HalcyonTM LA was performed successfully.

Details

Language :
English
ISSN :
23453672
Volume :
21
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Iranian Journal of Medical Physics
Publication Type :
Academic Journal
Accession number :
edsdoj.6a7efadc673498f9d9e982d8120e193
Document Type :
article
Full Text :
https://doi.org/10.22038/ijmp.2023.69197.2215