Comparação entre as medidas da espessura central corneana usando a paquimetria óptica e a ultra-sônica Comparison between the measurements of central corneal thickness using optic and ultrasonic pachymeters

Authors :
Aline Leonel Maimone
Nelson Maimone
Robson Marcelo Rossi
Source :
Revista Brasileira de Oftalmologia, Vol 66, Iss 5, Pp 309-314 (2007)
Publication Year :
2007
Publisher :
Sociedade Brasileira de Oftalmologia, 2007.

Abstract

OBJETIVOS: Comparar a medida da espessura corneana central (ECC) obtida pelo paquímetro óptico Haag-Streit e a paquimetria ultrassônica DGH 500 (PachetteTM). MÉTODOS: Foram avaliados, 200 olhos de 100 pacientes utilizando-se o paquímetro óptico (PO) e o ultrassônico (PU). As medidas foram realizadas na área central da córnea, respeitando a área dos 3 mm, em pacientes com córneas normais, em olhos hipermétropes, emétropes e míopes, excluindo doenças oculares, usuários de lentes de contato ou submetidos às cirurgias. RESULTADOS: A média geral da ECC medida pelo PO foi 603,8± 32,6µm, e 568,2±40,5 µm pelo PU. As comparações entre as medidas dos dois aparelhos foram realizadas ao nível de 5% de significância e a diferença entre os dois aparelhos foi 35,7±26,4 µm (p=0,0000), indicando diferença significativa entre os métodos utilizados. Não houve diferença estatística entre olhos hipermétropes, emétropes e míopes usando o PU. CONCLUSÃO: A medida da ECC é superestimada pelo PO quando comparada com o PU.PURPOSE: To compare measurements of central corneal thickness obtained using a Haag-Streit optic pachymeter and a DGH 500 (Pachette Ttm ) ultrasonic pachymeter in normal patients. METHODS: An evaluation was made of 200 eyes of 100 patients using Optic (PO) and Ultrasonic (PU) pachymeters. Measurements were made in the area of the central cornea (ECC) respecting the 3.0 mm territory, in patients with normal corneas of hypermetropic, emetropic and myopic eyes. Patients with ocular diseases, ocular surgeries, and contact lens wearers, were excluded. A statistical analysis was performed using a Paired Student's t test to compare measurements between instruments at the 5% level of significance. RESULTS: The mean thickness of the ECC measured by the PO was 603.8±32.6µm and by the PU, 568.2±40.5µm. The difference between the two instruments was 35.7±26.4µm. Applying the t test with p =

Details

Language :
English, Spanish; Castilian, Portuguese
ISSN :
00347280
Volume :
66
Issue :
5
Database :
Directory of Open Access Journals
Journal :
Revista Brasileira de Oftalmologia
Publication Type :
Academic Journal
Accession number :
edsdoj.5d4e1c972ee14763838a90859a367363
Document Type :
article
Full Text :
https://doi.org/10.1590/S0034-72802007000500004