Back to Search Start Over

In-vitro measurement of aerosol generated by Respimat SMI penetrating through upper airways

Authors :
Misik Ondrej
Lizal Frantisek
Malý Milan
Karas Jakub
Jedelsky Jan
Jicha Miroslav
Source :
EPJ Web of Conferences, Vol 269, p 01038 (2022)
Publication Year :
2022
Publisher :
EDP Sciences, 2022.

Abstract

Respimat soft mist inhaler (SMI) is an inhaler producing a soft mist with low velocity. The liquid drug solution is emitted by the mechanical energy of compressed spring, so the device is propellant-free. This paper describes the measurement of particle fraction penetrating through the model of the upper respiratory tract (mouth cavity, pharynx, larynx trachea). The solution inside the cartridge was marked by fluorescein in order to evaluate the regional particle deposition in the model. After the exposition, segments of the model, as well as the filter, were washed out and the deposited fluorescein was extracted into water. The analysis showed that the fraction of particles penetrating through the upper respiratory tract was 64.7±6.7 %. However, the real inhalation pattern was not simulated and measurement was provided with continual “inhalation” without “exhalation”.

Subjects

Subjects :
Physics
QC1-999

Details

Language :
English
ISSN :
2100014X
Volume :
269
Database :
Directory of Open Access Journals
Journal :
EPJ Web of Conferences
Publication Type :
Academic Journal
Accession number :
edsdoj.5a840648a9f14e94a44a074f940b55a6
Document Type :
article
Full Text :
https://doi.org/10.1051/epjconf/202226901038