Back to Search
Start Over
Establishment of ISO 4037-1 X-ray narrow-spectrum series
- Source :
- The Journal of Engineering (2019)
- Publication Year :
- 2019
- Publisher :
- Wiley, 2019.
-
Abstract
- In order to calibrate radiation protection instruments and determine radiation dose, it is critical to establish an X-ray narrow-spectrum series recommended by the standard ISO 4037-1 which is published by the International Standardisation Organisation (ISO). By experiment, the qualities of filtered X radiation of narrow-spectrum series in the range from 60 to 300 kV were established. The differences of the first half-value layer (HVL1) and the second HVL (HVL2) between experimental results and the values given in the standard ISO 4037-1 were all within 5%. What is more, the Monte Carlo code EGSnrc was used to simulate the spectra of these radiation qualities, which showed a good agreement with these spectra given in the standard ISO 4037-1. The spectral resolution, mean energy and effective energy of these reference qualities can be obtained, which are all have good agreement with the given values.
- Subjects :
- iso standards
monte carlo methods
calibration
dosimetry
radiation protection
voltage 60.0 kv to 300.0 kv
spectral resolution
monte carlo code egsnrc
first half-value layer
iso 4037-1 x-ray narrow-spectrum series
filtered x radiation
radiation dose
radiation protection instruments
radiation qualities
standard iso 4037-1
hvl2
Engineering (General). Civil engineering (General)
TA1-2040
Subjects
Details
- Language :
- English
- ISSN :
- 20513305
- Database :
- Directory of Open Access Journals
- Journal :
- The Journal of Engineering
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.5a68d5a1054c4118842c0fdec7b4eb78
- Document Type :
- article
- Full Text :
- https://doi.org/10.1049/joe.2018.9126