Three-channel ion chromatograph for improved metabolic evaluation of urolithiasis

Authors :
Qiang Li
Guanlin Liu
Yue Cheng
Wenbo Tang
Source :
BMC Urology, Vol 21, Iss 1, Pp 1-7 (2021)
Publication Year :
2021
Publisher :
BMC, 2021.

Abstract

Abstract Background Urolithiasis is a multi-etiological disease resulting from a combination of environmental and genetic factors. One of the most challenging aspects of this disease is its high recurrence rate. For most patients, an in-depth metabolic evaluation may reveal the presence of urinary stones. The fact that different urinary stone-related compounds (USRCs) are measured by different methods renders the metabolic evaluation of urolithiasis quite tedious and complex. Methods A three-channel ion chromatograph (IC) that automatically measures the concentration of common metabolic indicators of urolithiasis in urine (i.e., oxalate, citrate, uric acid, calcium, and magnesium) was developed to improve the efficiency. To validate its precision and specificity, standard curves were prepared using working solution of these indicators. 100 standard solutions of these indicators were measured with our new IC and three other ICs as the control instruments; analyte concentrations in 100 24-h urine samples from volunteers and 135 calculi patients were also measured. Results All analytes had good linear relationships in concentration ranges of 0–10 mg/L. The precision experiments in the standard and urine samples showed that the measurement errors of the newly developed IC were all less than 5%. In urine, the recovery rate ranged from 99.6 to 100.4%, the coefficient of variation ranged from 1.39 to 2.99%, and the results matched between our newly developed IC and the control ICs. The results of the efficiency test showed that we can finish the analysis at the average number of 14 people per day with the new IC. While the average number in the control group is 3.85/day (p = 0.000). Conclusions Overall, this multi-channel system significantly improves the efficiency of metabolic evaluation while retaining accuracy and precision.

Details

Language :
English
ISSN :
14712490
Volume :
21
Issue :
1
Database :
Directory of Open Access Journals
Journal :
BMC Urology
Publication Type :
Academic Journal
Accession number :
edsdoj.597dfa7258844cc496aedcd3309093df
Document Type :
article
Full Text :
https://doi.org/10.1186/s12894-021-00914-4