Back to Search Start Over

Development of an Arduino-based Control and Sensor System for a Robotic Laparoscopic Surgical Unit

Authors :
Fracisco Emmanuel Jr., III Munsayac
Nilo Bugtai
Renann Baldovino
Noelle Marie Espiritu
Lowell Nathaniel Singson
Source :
Recoletos Multidisciplinary Research Journal, Vol 12, Iss 1, Pp 125-144 (2024)
Publication Year :
2024
Publisher :
Center for Policy, Research and Development Studies, 2024.

Abstract

The LAPARA System, a Philippine-made robotic surgical system, tested its control system in this paper. Three types of tests are then done to the system: PID Optimization Test, Position Checking, and Data Transfer Rate and Memory Bandwidth Testing. Results from the PID resulted in the values 2.32 for the P, 0.4 for the I, and 1.5 for the D to be chosen to ensure the system runs smoothly. The system was also able to run properly during Position Checking, though movement in the Pitch and Yaw required refinement due to the constraints. Also, the data transfer rate for the PC to Arduino Due connection yielded a 128kb/s speed, slower than the 480 Mbps rating, while the memory bandwidth testing yielded results that allowed for storage of 23,040 32_bit values. In conclusion, although minor adjustments were needed to refine the system, the LAPARA system was able to perform as intended.

Details

Language :
English
ISSN :
24231398, 24083755, and 15088510
Volume :
12
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Recoletos Multidisciplinary Research Journal
Publication Type :
Academic Journal
Accession number :
edsdoj.553d150885104d989b4d2fa147056344
Document Type :
article
Full Text :
https://doi.org/10.32871/rmrj2412.01.10